They measure the output power of the electrodes to determine whether or not it is operating at the rated level. Other parameters such as the output current, output voltage, and crest factors are also checked to ensure that the machine is operating properly under various loads.

By testing the output on a variety of different load types, it is ensured that the unit responds to the various impedances that it encounters. In order to determine whether or not the device is functioning properly, it is necessary to consult the user manual or other sources for the standard values for various loads. It is acceptable for the output values to fall within a margin of +- 5% of the rated value; however, the lowest possible deviation is preferred over the highest possible deviation.

The vast majority of new electrosurgical generator unit  generator  units come equipped with built-in testing capabilities as a matter of standard equipment. Performing an output power test on your machine at a faster rate will be possible as a result of this. The machine is equipped with impedance circuitry, which allows it to handle the weight of the load. Alternatively, you can simply go through the testing sequence and look for the values to see if they are present. You'll have to experiment with different impedances between the electrodes if you don't know what you're doing.


At the moment, we are checking for leakage.

Being exposed to high-frequency alternating current can be extremely dangerous. When current couples capacitively with other conductive materials at high frequencies, a phenomenon known as capacitive coupling occurs. It is possible for both the operator and the patient to be harmed if there is a significant amount of leakage current present in the unit, according to the manufacturer. Failure to check for leakage currents frequently results in the appearance of burn marks on the surface. Therefore, it is critical to have the equipment tested before it is used so that it does not suffer from these types of damages.

Several international organizations, including the International electrosurgical generator unit units units Commission (IEC), have developed standards for the safe use of electrosurgical generator generator  units. According to these standards, the leakage current through a 200 ohms load to the ground should not be greater than 150 mA in any direction when the load is disconnected. These tests are carried out on all of the electrodes that are present in the device at any given time during the process of manufacturing it.

An electrode fault is introduced into the simulation and the electrodes are treated as if they were defective in order to simulate the device. The unit's response is then monitored, and the  generator unit  units unit must be protected from any type of leakage that exceeds the acceptable level of exposure.

Return Electrode Current Monitoring Test (Return Electrode Current Monitoring Test) is a test that measures the current flowing through an electrode.

Performing a monitoring test on the return electrode is also very important because it ensures that the device is functioning properly and helps to avoid any major mishaps during the procedure. Alarms are installed in  generator unit  units, and they can be of either an audio or visual nature depending on the situation. When the current flowing through the return electrode exceeds the threshold value set by the user, these alarms sound. It is necessary to turn off the generator in addition to sounding the alarm as soon as it is possible.

It is possible to complete this test with relative ease if you use standard impedances and increase the value for return electrode current. The majority of the time, the values for these standards are provided by the manufacturer in the user manual for the device in question.

Concluding Remarks

In the case of working with high-frequency alternating currents, it is absolutely critical that all of these tests be carried out prior to the operation being carried out in question. Any patient or member of the surgical team who is injured as a result of a malfunctioning  generator  generator  unit should seek medical attention immediately. In conjunction with the use of these tests, it is possible to ensure the safe operation of these devices.